2.3 Making Connections
51
3
2
Chapter 2 M
eas
ureme
n
t Prep
ar
at
ions
<Connecting to the terminals of the Wireless Universal Unit>
Required items: phillips screwdriver (No.2), flat-blade screwdriver (tip width 2.6 mm),
measurement leads
Recommended wire diameter:
Single strand diameter:
0.4 mm -
1.2 mm (AWG26-16)
Multi-strand:
0.2 mm
2
- 0.75 mm
2
(AWG24-20)
Standard insulation stripping length: 8 mm
Resistance Measurement
1
With a Phillips screwdriver,
loosen the screws in the terminal
block cover, and lift the cover.
2
Push in the button on the con-
nector with a flatblade screw-
driver.
3
With the button held in,
insert the cable into the cable
connection hole.
6
Replace the terminal block
cover, and tighten the screws.
Tighten the terminal block cover
screws to 0.6 N
m or less.
4
Release the button.
The cable is locked.
4-wire configuration
(2- and 3-wire configurations are not
supported)
5
Connect to the measurement
object.
2
3
4
button
5
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......