5.1 Triggering Measurement Start and Stop
133
3
5
Chapter 5 S
p
ecif
ying Crite
ria
for M
easurement
s
Using a Window Trigger
Set the range within which you wish to start or stop measure-
ment using the upper and lower limit values.You can select
whether measurement starts or stops when the input signal level
enters (In) or exits (Out) of this window.
Upper and lower thresholds can be verified in the monitor or on
the Waveform screen. When scaling is enabled, the scaled
threshold values are displayed.
Make trigger settings on the
[CH]
screen of the displayed chan-
nel.
See:
"Key Setting Procedure" (p .130)
T
T
T
T
IN
OUT
1
Select
[Window]
as the start/stop trigger type.
2
Select trigger timing (window type).
3
Enter the upper/lower threshold value.
Setting options: (
: default setting)
In
Trigger when the input signal level enters the
window defined by the upper/lower thresholds.
Out
Trigger when the input signal level exits the win-
dow defined by the upper/lower thresholds.
1
2
3
3
Scaled values are displayed
when scaling is enabled.
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......