3.7 Scaling Settings (as needed)
95
3
3
Chapt
er
3 Sett
ings
Use the scaling function to convert input volt-
age to the physical units of the measurement
parameter for display, such to convert voltage
input for display as electrical current.
Converted values can be displayed in fixed or
floating-point notation.
Make these settings on the
[CH]
screen.
See:
"Key Setting Procedure" (p .91)
3.7 Scaling Settings (as needed)
Scaling Voltage, Temperature, Humidity, Resistance, Revolution, and
Current Measurements
Normal Display
(Off)
Scaling Enabled
A
mV
The Setting screen cannot be
displayed while measuring.
1
Select the desired numerical notation.
2
Select the scaling method.
:
3
Enter the
[Unit]
name on the comment entry
screen. (Up to 7 characters)
See:
About Comments (p. 98)
4
Ratio: Enter the conversion ratio and offset.
2-pt: Enter the values at two points before and
after conversion.
Move the cursor to the numerical value and press
EN-
TER
to display the numerical setting dialog.
See:
"Numerical Value Entry" (p .93)
Setting options:(
: default setting)
Dec
Displays converted values using fixed-point
notation.
Exp
Displays converted values using floating-
point notation.
Off
Scaling is disabled.
Setting options:(
: default setting)
Ratio
Convert by specifying the ratio of physical
units per volt of input signal (conversion ra-
tio), offset, and the name of the units.
2-pt
Convert by specifying the input signal volt-
age at two points, the physical unit values at
those points, and the name of the units.
1 2
3
Conversion Ratio Setting
2-Point Setting
2-Point Setting
With
[Cond]
set to
[Exp]
With
[Cond]
set to
[Dec]
4
Conversion Ratio Setting
A value measured on CH1 of UNIT1 (Analog channel CH1-1) can be set as the scaling
value (Inter-Channel Compensation function) (p. 107).
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......