5.6 Canceling a Schedule
146
About operation restrictions while
[Reserved]
is displayed
(standby state)
• Keys related to starting, stopping, files, and saving do not work.
• To change a setting, first cancel the schedule. If you change a setting in the
standby state, a dialog box to confirm canceling the schedule operation will
appear.
About schedule start and stop operations
• Normal settings files cannot be saved or loaded as schedule settings. Select
schedule settings to save on the Schedule Confirmation screen or when sav-
ing selections and load the corresponding file as necessary.
• If the first schedule start time has passed at the point in time that the schedule
is set, the set measurement conditions are accessed and measurement starts
immediately.
• "The reservation will be started soon." is displayed 1 minute before a sched-
ule start time, and then measurement starts at the start time.
• When repeat recording is set to
[Off]
, the schedule operation finishes when
one measurement ends. When repeat recording is set to
[On]
, measurement
is performed repeatedly until the stop time.
• When the state is
[Waiting for pre-trigger]
or
[Waiting for trigger]
at the
stop time, measurement stops and the state switches to the waiting for next
schedule (schedule standby) state. When the state is
[Storing]
, measure-
ment is halted, calculation, and saving is performed using the data up until
that point in time, and then the state switches to the next schedule standby
state.
• If the next schedule start time has passed because of the time required for
saving, the measurement conditions are accessed and the next measure-
ment starts immediately.
• When all schedules are complete, the scheduled state is cancelled.
• If the power is cut off when measurement is scheduled, the scheduled state is
resumed when the power recovers and measurement is started if the time
when the power recovers is between the schedule start time and schedule
stop time, regardless of the Start Backup setting (On/Off).
• When schedule settings include errors, for example due to overlap with a
properly configured schedule, only properly configured schedules will be exe-
cuted.
• Lower schedule numbers take precedence when checking for overlapping
schedule setting times.
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......