5.1 Triggering Measurement Start and Stop
138
You can apply a trigger at a fixed time interval. This can be set on the
[Trig&Alm]
screen.
Applying Trigger at Fixed Time Interval (Interval Trigger)
1
Select the
[Trig&Alm]
Screen.
2
Enable triggering
[On]
.
3
Select the interval conditions.
4
Set the time interval for repeat measurement.
[Day]/ [Hour]/ [Min]/ [Sec]
A trigger is applied simultaneously with the start of mea-
surement, and then a trigger is subsequently applied at
the specified measurement interval.
Setting options: (
: default setting)
O
ff
Disables the interval trigger.
OR
Applies the interval trigger using OR logic
(see figure below).
AND
Applies the interval trigger using AND logic
(see figure below).
2
4
Displays the maximum
amount of time that can be set.
Change 1's digit
Change 10's digit
When using the Logger Utility to per-
form measurement, interval triggers are
disabled.
3
Trigger
criteria
OR
AND
T
:Trigger point
:Other trigger criteria
Interval
Interval
T
Measurement start
T
T
No trigger event occurs.
Interval
Interval
Measurement start
T
One trigger is valid during the interval.
T
No trigger event occurs if criteria do not match other trigger criteria.
(OR logic is used if no other trigger criteria have been set.)
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......