5.3 Confirming All Trigger and Alarm Criteria Settings
140
You can view and change trigger and alarm criteria settings for all channels on the
[Trig & Alm]
screen.
5.3 Confirming All Trigger and Alarm Criteria
Settings
Press
ENTER
to display the dia-
log at the right for the cursor-se-
lected channel. Trigger and
alarm criteria for the selected
channel can be set.
Trig Func (p. 135)
Select On to enable triggering.
Select
[Batch Proc]
to initialize settings and to copy trigger and alarm settings from channels
1-1, 2-1, 3-1, 4-1, 5-1, 6-1, 7-1 and W1 to W16 to any other channel. Copied settings include
level, slope, and threshold values.
See:
"3.10 Viewing and Editing with the All-Channel Settings List" (p. 101)
Input type
: Voltage
: Temperature (Thermocouple)
: Temperature (RTD)
: Humidity
: Resistance
:
Integration
:
Rotation
:
Logic
:
Current
: Alarm
:
Temperature
: Fungal index
: Fungal growth projection
Channel-specific settings (start, stop and alarm
triggers) are displayed to the right of the list for
the channel selected (highlighted) by the cursor.
Alarm
Select On to enable alarms.
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......