5.2 Trigger Setting Examples
139
3
5
Chapter 5 S
p
ecif
ying Crite
ria
for M
easurement
s
Following are examples of typical trigger settings.
5.2 Trigger Setting Examples
Intended Measurement Objective
Ref. No.
(next table)
Acquire data from when you press
START
until you press
STOP
No.1
Acquire data for one minute after each time you press
START
No.2
Acquire data at one-minute intervals for sixty minutes after you press
START
No.3
When the temperature measured on CH1 exceeds 500
C, acquire data until you press
STOP
.
No.4
Acquire data from when you press
START
until the temperature measured on CH1 exceeds
500
C.
No.5
Acquire data from when the temperature measured on CH1 exceeds 500
C until it drops below
300
C.
No.6
Acquire data whenever the temperature measured on CH1 exceeds 500
C, until it drops below
300
C.
No.7
Acquire data for only one minute from when the temperature measured on CH1 exceeds
500
C.
No.8
Acquire data for one minute before and after the time when the temperature measured on CH1
exceeds 500
C.
No.9
No.
[Setting]
Screen
[Trig & Alm]
Screen
[CH]
Screen
Repeat Record time
Split
Save
Trig
Func Timing
Trig
Source Pre-Trig
Start Trig.
Stop Trig.
1
Off
[Cont]
Off
Off
-
-
-
-
-
2
Off
[Time]
0h: 1min: 0s
Off
Off
-
-
-
-
-
3
Off
[Time]
1h: 0min: 0s
On
[Split
Length]
1 min
Off
-
-
-
-
-
4
Off
[Cont]
Off
On
[Start] [Start]
OR
-
[Cond]Level
[Level]
500
C
-
5
Off
[Cont]
Off
On
[Stop] [Stop]
OR
-
-
[Cond]Level
[Level]
500
C
6
Off
[Cont]
Off
On
[Start/
Stop]
[Start]
OR
[Stop]
OR
-
[Cond]Level
[Level]
500
C
[Cond]Level
[Level]
300
C
7
On
[Cont]
Off
On
[Start/
Stop]
[Start]
OR
[Stop]
OR
-
[Cond]Level
[Level]
500
C
[Cond]Level
[Level]
300
C
8
Off
[Time]
0h: 1min: 0s
Off
On
[Start] [Start]
OR
-
[Cond]Level
[Level]
500
C
-
9
Off
[Time]
0h: 2min: 0s
Off
On
[Start] [Start]
OR
1 m
[Cond]Level
[Level]
500
C
-
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......