9.1 External Trigger Input
205
3
Chapter 9 Ext
ernal Cont
rol
9
The external control terminals on the instrument support trigger signal input and
output.
Be sure to read p. 52 to p. 53 for external control terminal connection details.
Triggering can be controlled by applying a signal from an external trigger source
(p. 136).
This allows synchronous operation of multiple instruments by parallel triggering
See "Using External Triggering"(p .136) for setting procedures and trigger signal
details.
The external signal input function can be changed to insert event marks.
See:
"Inserting Event Marks Using External Input Signals"(p .122)
External Control
Chapter 9
9.1 External Trigger Input
External signal
input
(EXT.TRIG)
Trigger event occurs.
2.5 ms or
greater
2.5 ms or greater
HIGH: 2.5 to 5.0 V
LOW: 0 to 0.8 V
Triggering occurs at the selecting rising or falling edge.
(according to the selected
[Ext trig]
detection edge)
Voltage range
HIGH level: 2.5 to 5.0 V, LOW level: 0 to 0.8 V
Pulse width
HIGH period: 2.5 ms or greater, LOW period 2.5 ms or greater
Maximum input voltage
DC0 to 10 V
Trigger Input Signals
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......