3.3 Input Channel Settings
74
Level monitor numerical value refresh rate and waveform horizontal axis span
depend on the recording interval and horizontal axis (time base) settings.
Therefore, long recording intervals can delay numerical value display
refresh by as much as 5 seconds. If you need faster value display refresh,
select a shorter recording interval.
If the horizontal axis (time span) is too long, one division of the waveform
may not be displayed.
Monitor display switching
Allows you to switch the information dis-
played by the level monitor.
(Settings on left)
[Wave]
Displays the waveform.
[Value]
Displays values for the units set on the
right side of the screen.
(Settings on right)
Allows you to select which waveform to
display if the setting on the left includes
any
[wave]
.
[All Wave]
Displays the waveforms for all channels.
[UnitWave]
Displays the waveforms for all channels
from the units set on the right side of the
screen.
[Ch Wave]
Displays the waveforms for channels set
on the right side of the screen.
Level Monitor
Selecting
[Wave]
causes the scale for the
currently selected channel to be displayed.
The gauge scale is determined by measure-
ment and display range settings.
Monitoring is not available for disabled
(Measurement Off) channels, and those
with
[X]
selected for their color.
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......