3.2 Configuring Measurement Settings
69
3
3
Chapt
er
3 Sett
ings
Recording Interval (Interval)
Select the data acquisition interval.
Repeat Recording (Repeat)
Select whether to automatically start a new measurement after stopping, or to finish after
one measurement.
Recording Time (Record time)
Select whether to measure for a specified time span, or continuously.
Recording time specification(Record time specification)
Set when the record time is set to
[Time]
.
Specify the duration of one recording length (time span).
The maximum setting is limited by the recording capacity of internal buffer memory. The
actual limit depends on the selected recording interval and number of channels used for
measurements, so the maximum settable time is displayed as the
[Max]
value (indicated by
the red box).
Setting options:(
: default setting)
100 ms
, 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 1 min, 2 min, 5 min, 10 min, 20 min, 30
min, 1 h
Setting options:(
: default setting)
Off
Measuring stop after one recording length (time span).
On
When stopped by a stop trigger or after measurement for the specified amount of
time, measurement automatically restarts. Press
STOP
to stop measuring.
Setting options:(
: default setting)
Time
Use when you wish to specify the recording time for measurement.
Cont
Use when you wish to conduct measurement until the
STOP
key is
pressed or until the strop trigger criteria are satisfied.
Setting options:(default setting: 10 seconds)
d (0 to 499), h (0 to 23), min (0 to 59), s (0 to 59)
[Repeat]
On
Off
Change 1's digit
Change 10's digits
[Max]
,
[Data pts]
If you wish to record to the SD memory card or USB memory stick beyond the maximum recording time
for the instrument’s internal memory, set the record time to
[Cont]
.
"Maximum Recordable Time" (p .14)
You cannot specify 0 d 0 h 0 min 0 s as the time. If you do so, the time will automatically be set to 0 d
0 h 0 min 1 s.
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......