12.1 LR8410-20 Wireless Logging Station Specifications
277
3
Chapter 12
S
p
ec
ifica
tions
12
USB Function
• Data acquisition and measurement criteria setting with the Logger Util-
ity
• Setting and measurement by communications commands
• USB Drive Mode
(Data can be sent from an external SD memory card to a computer.)
Communications Functions
Display Section
Display type
5.7-in TFT Color LCD (640 x 480 dots)
Display resolution
(with Waveform display
selected)
Max. 16 divisions (time axis) × 10 divisions (vertical axis)
(1 division = 36 dots (time axis) × 36 dots (vertical axis)
Display character
English/Japanese selectable
Backlight life span
Approx. 50,000 hours (@23°C, 73.4°F)
Backlight saver
The backlight turns off when no key is pressed
(Off, or 1 to 30 minutes)
Backlight Brightness
Selectable from four levels (100%, 70%, 40%, 25%)
Power source
AC adapter
Z1008 AC Adapter(DC12 V)
Rated supply voltage
100 V to 240 V AC
Rated supply
frequency
50 Hz/60 Hz
Battery
Z1007 Battery Pack
(The AC Adapter has priority when connected)
Li-ion 7.2 V 2170mAh
External Power
10 V to 28 V DC
Power consumption
Typical power
consumption
With Z1008 AC Adapter or other 12 V DC external power, without battery
pack
w/LCD max. brightness: 8 VA (LR8410 only)
w/LCD backlight off: 7 VA (LR8410 only)
Maximum rated
power
When using the Z1008 AC Adapter:
45 VA (including the AC adapter), 15 VA (LR8410 only)
When using an external DC28 V power source:
15 VA (when battery is charged, and w/LCD max. blightness)
When using a Z1007 Battery Pack :7 VA (w/LCD max. brightness)
Continuous operating
time
When using Model Z1007 Battery Pack, approx. 3 hours
(Backlight brightness 25%, @23°C, 73.4°F)
Charging function
The AC adapter or a 10 V to 28 V DC external power supply can be con-
nected while the Z1007 Battery Pack is installed.
Charging time: Approx. 7 hours (@23°C, 73.4°F)
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......