10.2 LAN Settings and Connections (Before connecting to the network)
233
3
Chapter 10
Connec
tion t
o
a
PC (
C
ommunic
ation)
10
Confirm cable connections.
When connecting one instrument to one PC, use the supplied cross-over adapter with the 9642
LAN Cable.
In case of poor connection, disconnect and reconnect the cable to clean the contacts. When
connecting the cable, the green LED on the instrument's LAN jack should light.
Confirm the PC's IP address.
Click
[Run...]
on the Windows
®
Start menu (or select
[All Programs]
-
[Accessories]
-
[Comand Prompt]
) and enter “
ipconfig /all
” to confirm the IP address of the network interface,
subnet mask, and gateway address.
If settings are incorrect:
Refer to "1. PC Network Setup"(p. 226) to reset the IP address.
Confirm communication between the instrument and PC.
After confirming that the IP addresses of the instrument and PC are correct, use the “
ping
” util-
ity to test data transfer.
1.From the Start menu, select
[All Programs]
-
[Accessories]
-
[Comand Prompt]
.
2.(at the blinking cursor)
Enter "
ping XXX.XXX.X.X
"(the IP address of the host device to test).
Alternatively, the host name can be entered if its IP address is assigned by DNS.
For example, if the instrument's IP address is 192.168.1.2, enter "
ping 192.168.1.2
" and
press Enter.
The following display text indicates normal operation. The
“
time
”
is the communication
response time.
Pinging 192.168.1.2 with 32 bytes of data:
Reply from 192.168.1.2: bytes=32 time<10ms TTL=32
Reply from 192.168.1.2: bytes=32 time<10ms TTL=32
Reply from 192.168.1.2: bytes=32 time<10ms TTL=32
Reply from 192.168.1.2: bytes=32 time=1ms TTL=32
The following display text indicates communications failure.
Recheck cable connections.
Pinging 192.168.1.2 with 32 bytes of data:
Reply from 192.168.1.2: Host is down.
Reply from 192.168.1.2: Host is down.
Reply from 192.168.1.2: Host is down.
Reply from 192.168.1.2: Host is down.
When LAN Communication Fails
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......