5.4 Starting & Stopping Measurement on a Specified Day
141
3
5
Chapter 5 S
p
ecif
ying Crite
ria
for M
easurement
s
You can specify dates as desired for measurement to start and stop. Before config-
uring these settings, ensure that the instrument’s clock is set to the correct time.
Before setting, confirm that the instrument clock is set to the correct time. If not,
reset the clock on the
[System]
screen (p. 202).Instrument settings cannot be
changed once these settings are active. Configure instrument settings before
scheduling measurement.
5.4 Starting & Stopping Measurement on a
Specified Day
1
Select
[Res Set.]
2
Select the checkbox [
] for the number (No. 1
to No. 10) of the schedule you wish to set.
3
Set
[Spc.Dt (Specified Date)]
in
[Type]
for the
schedule.
4
Set the date and time you want to start and stop
measurement.
Set each of year-month-day, hour, and min-
ute.You can enter the current time by selecting
the clock mark
and pressing the
ENTER
key.
Setting options: (
: default setting)
Off*
Disables the schedule with the selected number.
Spc.Dt
Select when you wish to start and stop mea-
surement by specifying a date and time.
Reg
Select when you wish to start and stop mea-
surement regularly, for example every day or
on specific days.
1
1.Schedule the measurement.
Select the
[Setting]
screen.
(continues on next page)
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......