4.1 Confirming Measured Values, and Starting Measurement
111
3
Chapt
er
4 Observing
Me
asurement
s and
D
at
a
4
Before starting a formal measurement, press
MONITOR
to check measured values
as needed (p. 40).
When you have confirmed that the settings are correct, press
START
to start formal
measurement (p. 36).
Pay attention to the following items while measuring.
• If a power outage affects only measurement units or wireless loggers
Data during the outage will be treated as lost. Measurement can be continued once
power is restored.
• If the outage is limited to the instrument
Set the smart backup setting to on. Recording will start automatically when power is
restored. Since measurement data from before the outage will not be stored in the instru-
ment’s internal buffer, it is recommended to automatically save data to the included or
optional SD memory card. Data will be lost while the power to the instrument is inter-
rupted, even if measurement units and wireless loggers continue to receive power.
Both measurement units and wireless loggers have internal backup memory.
When communications are restored, measurement units will retransmit (65,536/n) data
points (where n indicates the number of channels) to restore the data on the Wireless
Logging Station.
If there are multiple units to which communications are restored, data will be restored in
the order in which the units are reconnected. For example, a communications failure can
last for about 72 minutes without loss of data when measuring 15 channels of data at a
recording interval of 1 sec. with a measurement unit.
See:
"When Temporarily Unable to Communicate with a Measurement Unit" (p. A34)
Observing Measurements and
Data
Chapter 4
4.1 Confirming Measured Values, and Starting
Measurement
If the power goes out during measurement
If communications fail temporarily
Measure
(as needed)
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......