Appendix 12 Instrument/Unit Communications Interruptions and Reestablishment
A
38
If communications with a unit are interrupted so that there is no data for a channel tar-
geted by a trigger, no trigger judgment will be performed for that channel. Trigger judg-
ment is only performed for channels for which there is data available. Once
communications have been reestablished with the unit and data has been restored from
the unit, trigger judgment will be performed for restored data all at once, and trigger
events will occur as trigger criteria are satisfied. Trigger output will be generated as trig-
ger judgments are made. For stop triggers, data after the stop trigger criteria are satisfied
will be reported as “NO DATA.”
Triggers
When using a pre-trigger, the trigger will not activate for restored data.
Additionally, data will not be restored while waiting for a pre-trigger event.