12.1 LR8410-20 Wireless Logging Station Specifications
278
Trigger Function
Trigger method
Digital comparison
Trigger modes
Single or continuous
Trigger timing
Start, stop, start&stop
Trigger source
Analog, pulse and
logic input
Up to 105 channels, depending on how many LR8510 Wireless Voltage/
Temp Unit, LR8511 Wireless Universal Unit, LR8512 Wireless Pulse
Logger, LR8513 Wireless Clamp Logger, LR8514 Wireless Humidity
Logger, LR8515 Wireless Voltage/Temp Logger and LR8520 Wireless
Fungal Logger are connected (U1-1 to U7-15).
External trigger
Trigger on external input signal rising or falling edge (selectable).
Interval trigger
The trigger is activated at a measurement interval set in days, hours,
minutes, and seconds.
Trigger criteria
• AND or OR of each trigger source
• Trigger criteria can be set for each channel.
• Free-running when all are Off
Trigger types
Level Trigger
Triggering occurs when the signal rises (or falls) through a specified
value.
Window Trigger
Triggering occurs when the input signal level enters or exits the ampli-
tude window defined by upper/lower thresholds.
Pattern trigger
Applies the trigger when a pattern defined in terms of 1, 0, and values
is matched (where " " indicates either value).
*Setting only available when using logic measurement with the LR8512.
Trigger response time
Recording in 3 sec. or less
(with 1 unit, good communications)
Recording in 5 sec. or less
(When seven units are connected, good communications)
Trigger level resolution
Analog: 0.1% f.s. (f.s. = 10 frames)
Pulse: Integration 1c, RPM 1/n (where n equals the pulse count setting
per revolution)
*Pulse setting only available when using the LR8512.
Rising
: triggers upon transition from 0 V to 0.8 V (LOW)
to 2.5 V to 5.0 V (HIGH).
Falling
: triggers upon transition from 2.5 V to 5.0 V
(HIGH) to 0 V to 0.8 V (LOW), or when the exter-
nal trigger terminal is shorted to a GND terminal.
Trigger input voltage
range
: 0 V to 10 V DC
External trigger filter : On/Off
Minimum response
pulse width
: H period 2.5 ms or greater, L period 2.5 ms or
greater
Trigger input
terminals
: Push-button type terminal block
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......