9.3 Alarm Signal Output (Alarm Output)
207
3
Chapter 9 Ext
ernal Cont
rol
9
This signal is output when alarm criteria are satisfied.
Specify the desired alarm criteria.
See:
9.3 Alarm Signal Output (Alarm Output)
Alarm signal output
(ALM1 to ALM4)
When alarm criteria are satisfied
The alarm LED will light up (turn red) when
any of the ALM1 to ALM4 alarm criteria are
satisfied.
About the Alarm Output Signal
Output signal
Open drain output (with 5 V voltage output), active LOW
Output voltage
level
HIGH level: 4.0 to 5.0 V, LOW level: 0 to 0.5 V
Output
response time
Recording in 3 sec. or less
(When one measurement unit or wireless logger is connected, good com-
munications)
Recording in 5 sec. or less
(When seven measurement units or wireless loggers are connected,
good communications)
When the communications state is poor, it may take some time for the
alarm to be output.
Maximum switch
rating
5 to 30 VDC, 200 mA