13.2 Troubleshooting
302
If the cause is unknown
A waveform does
not appear when
you press
START
.
• Is the "Waiting for pre-trigger" message
displayed?
• Is the "Waiting for trigger" message dis-
played?
• Is the waveform display color selection box
selected (
)?
When pre-triggering is en-
abled, triggering is ignored un-
til the pre-trigger portion of the
waveform has been acquired.
Recording starts when a trig-
ger occurs.
No changes occur
in the displayed
waveform.
• Is the measurement cable connected cor-
rectly?
• Is the measurement range set properly?
Verify that the measurement
cable or connection cable is
connected correctly (p. 46).
Data cannot be
saved to the
removal storage.
• Is the removal storage inserted properly?
• Is the removal storage formatted?
• Is the remaining capacity of the removal
storage too low?
"Using a SD memory card/
USB flash drive" (p. 60)
Z1007 battery pack
cannot be charged
(Charging LED is
not lit).
• Please confirm that the surrounding tem-
perature is within 5 to 35°C range.
The temperature range within
which the instrument and
measurement unit can be
used to charge the battery
pack is 5°C to 35°C.(p. 10)
• Is the instrument and measurement units
stored for a long time in a plugged condi-
tion?
The battery pack may have
deteriorated and the battery
life may be expiring soon.
Please purchase a new bat-
tery pack. Please contact your
authorized Hioki distributor or
reseller.If the instrument is not
used for more than a month,
please remove the battery
pack for storage.(p. 10)
The battery back
operating time is
dramatically lower
than before,
even immediately
after charging
completes.
• Possibility of capacity decrease due to the
deterioration of the battery pack.
The battery pack may have
deteriorated and the battery
life may be expiring soon.
(p. 299)
Please purchase a new bat-
tery pack.
Please contact your autho-
rized Hioki distributor or re-
seller.
Try performing a system reset (p. 203).
All settings are returned to their factory defaults(p. A12).
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......