3.3 Input Channel Settings
79
3
3
Chapt
er
3 Sett
ings
Enable this channel setting for temperature measurement with the LR8514 Wireless
Humidity Logger, R8520 Wireless Fungal Logger. Make these settings on the
[CH]
screen.
See:
"Key Setting Procedure" (p .73)
Temperature Measurement Settings (When LR8514 or LR8520 is
used)
1
Select the input unit (
UNIT1
to
7
) and channel
(
1-1
to
7-15
), and check the box to enable the
channel.
The measurement setting applies to the two
temperature and humidity inputs for the
LR8514 and to the four temperature, humidity,
fungal index, and fungal growth projection
inputs for the LR8520.
2
The input type is fixed to temperature, and the
measurement range is fixed to 100°C.
-
100
°C
to
100
°C
: On
: Off
1
*
The range is fixed at
[100% f.s.]
.
2
*
When a measured value far exceeds the range limit (range label)
"OVER" or "-OVER" is displayed in place of numerical/cursor values, and the following values
are applied to calculations and data storage.
Input type
Input range +OVER value -OVER value
Temperature
100
°C
3276.7
-3276.8
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......