5.1 Triggering Measurement Start and Stop
136
To use an external signal as a trigger source, make the following settings.
Using External Triggering
1
Connect the external trigger signal to the
instrument's EXT.TRIG external control ter-
minal.
(See the connection procedure on P.53 )
2
Press
SET
to open the
[System]
screen.
3
Set the
[External Trig In]
to
[Trigger]
(default
setting).
4
Press
SET
to open the
[Trig & Alm]
screen.
5
Set triggering criteria (p. 135).
6
Select the slope for the
[Ext trig] [Start]
and
[Stop]
settings.
Select whether triggering occurs on the trigger sig-
nal's upslope or downslope.
This setting is forced Off when
[External Trig In]
is
set to
[Event]
on the
[System]
screen.
7
Confirm whether triggering is provided by an
external signal or by shorting the EXT.TRIG
terminal to GND (only when
[Ext trig]
is set to
[
]
).
Setting options: (
: default setting)
Off
External triggering is disabled.
Triggering occurs on the upslope from Low
(0 to 0.8 V) to High (2.5 to 5.0 V).
Triggering occurs on the downslope from
High (2.5 to 5.0 V) to Low (0 to 0.8 V), or
when the EXT TRIG terminal is connected
to a GND terminal.
3
2
4
6 6
Содержание LR8410-20
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents x...
Страница 32: ...Operating Precautions 20...
Страница 78: ...2 7 Compensating for Input Circuit Offset Zero Adjustment 66...
Страница 122: ...3 10 Viewing and Editing with the All Channel Settings List 110...
Страница 138: ...4 3 Marking Waveforms and Searching Marks Search Function 126...
Страница 222: ...9 4 Synchronous Measurements with Multiple instruments 210...
Страница 316: ...13 3 Disposing of the instrument 304...
Страница 360: ...Index Index 4...
Страница 361: ......
Страница 362: ......
Страница 363: ......
Страница 364: ......