Integrating into a ring topology
The device is integrated via the terminal connection for signal transmission (output 1) and
the connection to the service interface (CDI-RJ45).
Note the following when connecting:
• Recommended cable: CAT5e, CAT6 or CAT7, with shielded connector (e.g. brand:
YAMAICHI ; Part No Y-ConProfixPlug63 / Prod. ID: 82-006660)
• Maximum cable thickness: 6 mm
• Length of connector including bend protection: 42 mm
• Bending radius: 2.5 x cable thickness
N
ic
h
t
u
n
te
r
ar
e
ö
ffn
e
n
Nicht unter
Spannung öffnen
Do not open when
energized
Ne pas ouvrir
sous tension
Power
≤
42 (1.6)
1
2
A0033717
1
PROFINET connection
2
Service interface (CDI-RJ45)
An adapter for RJ45 and the M12 connector is optionally available:
Order code for "Accessories", option
NB
: "Adapter RJ45 M12 (service interface)"
The adapter connects the service interface (CDI-RJ45) to an M12 connector mounted
in the cable entry. Therefore the connection to the service interface can be established
via an M12 connector without opening the device.
7.4
Ensuring potential equalization
7.4.1
Requirements
L
CAUTION
Electrode damage can result in the complete failure of the device!
‣
Same electrical potential for the fluid and sensor
‣
Company-internal grounding concepts
‣
Pipe material and grounding