Electrical connection
Proline Promag W 500 PROFINET
50
Hauser
7.2
Connecting the measuring device: Proline 500 -
digital
NOTICE
Limitation of electrical safety due to incorrect connection!
‣
Have electrical connection work carried out by appropriately trained specialists only.
‣
Observe applicable federal/national installation codes and regulations.
‣
Comply with local workplace safety regulations.
‣
Always connect the protective ground cable before connecting additional cables.
‣
For use in potentially explosive atmospheres, observe the information in the device-
specific Ex documentation.
7.2.1
Connecting the connecting cable
L
WARNING
Risk of damaging the electronic components!
‣
Connect the sensor and transmitter to the same potential equalization.
‣
Only connect the sensor to a transmitter with the same serial number.
‣
Ground the connection housing of the sensor via the external screw terminal.
Connecting cable terminal assignment
+
–
A
B
61 62 63 64
+
–
A
B
61 62 63 64
4
2
1
4
3
5
6
A0028198
1
Cable entry for cable on transmitter housing
2
Protective earth (PE)
3
Connecting cable ISEM communication
4
Grounding via ground connection; on device plug versions grounding is through the plug itself
5
Cable entry for cable or connection of device plug on sensor connection housing
6
Protective earth (PE)
Connecting the connecting cable to the sensor connection housing
Connection via terminals with order code for "Sensor connection housing":
Option
A
Connecting the connecting cable to the transmitter
The cable is connected to the transmitter via terminals → 52.
Содержание Proline Promag W 500
Страница 262: ...www addresses endress com...