Electrical connection
Proline Promag W 500 PROFINET
66
Hauser
7.4.2
Connection example, standard scenario
Metal, grounded pipe
A0016315
19
Potential equalization via measuring tube
7.4.3
Connection example in special situations
Unlined and ungrounded metal pipe
This connection method also applies in situations where:
• The customary potential equalization is not used
• Equalizing currents are present
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)
DN
300
≤
DN
350
≥
A0029338
20
Potential equalization via ground terminal and pipe flanges
1. Connect both sensor flanges to the pipe flange via a ground cable and ground them.
2. If DN ≤ 300 (12"): Mount the ground cable directly on the conductive flange coating
of the sensor with the flange screws.
3. If DN ≥ 350 (14"): Mount the ground cable directly on the metal transport bracket.
Observe screw tightening torques: see the Sensor Brief Operating Instructions.
4. Connect the connection housing of the transmitter or sensor to ground potential by
means of the ground terminal provided for the purpose.
Plastic pipe or pipe with insulating liner
This connection method also applies in situations where:
• The customary potential equalization is not used
• Equalizing currents are present
Ground cable
Copper wire, at least 6 mm
2
(0.0093 in
2
)