Technical data
Proline Promag W 500 PROFINET
228
Hauser
Pulse/frequency/switch output
Function
Can be set to pulse, frequency or switch output
Version
Open collector
Can be set to:
• Active
• Passive
• Passive NAMUR
Ex-i, passive
Maximum input values
DC 30 V, 250 mA (passive)
Open-circuit voltage
DC 28.8 V (active)
Voltage drop
For 22.5 mA: ≤ DC 2 V
Pulse output
Maximum input values
DC 30 V, 250 mA (passive)
Maximum output current
22.5 mA (active)
Open-circuit voltage
DC 28.8 V (active)
Pulse width
Configurable: 0.05 to 2
000 ms
Maximum pulse rate
10
000 Impulse/s
Pulse value
Adjustable
Assignable measured
variables
• Volume flow
• Mass flow
• Corrected volume flow
Frequency output
Maximum input values
DC 30 V, 250 mA (passive)
Maximum output current
22.5 mA (active)
Open-circuit voltage
DC 28.8 V (active)
Output frequency
Adjustable: end value frequency 2 to 10
000 Hz (f
max
= 12
500 Hz)
Damping
Configurable: 0 to 999 s
Pulse/pause ratio
1:1
Assignable measured
variables
• Volume flow
• Mass flow
• Corrected volume flow
• Flow velocity
• Conductivity
• Electronics temperature
Switch output
Maximum input values
DC 30 V, 250 mA (passive)
Open-circuit voltage
DC 28.8 V (active)
Switching behavior
Binary, conductive or non-conductive
Switching delay
Configurable: 0 to 100 s
Содержание Proline Promag W 500
Страница 262: ...www addresses endress com...