Diagnostics and troubleshooting
Proline Promag W 500 PROFINET
174
Hauser
LED
Color
Meaning
Flashing red
A diagnostic event with "Warning" diagnostic
behavior has occurred.
Red
A diagnostic event with "Alarm" diagnostic behavior
has occurred.
Flashing red/green
The device restarts.
2
Device status (during
start-up)
Flashes red slowly
If > 30 seconds: problem with the boot loader.
Flashes red quickly
If > 30 seconds: compatibility problem when
reading the firmware.
3
Flashing/network status Green
Cyclic data exchange is active.
Flashing green
Following request from automation system:
Flash frequency: 1 Hz (flash functionality: 500 ms
on, 500 ms off)
Cyclic data exchange is not active, no IP address is
available:
Flash frequency: 3 Hz
Red
IP address is available but there is no connection to
the automation system.
Flashing red
Cyclic data exchange was active but the connection
was disconnected:
Flash frequency: 3 Hz
4
Port 1 active:
PROFINET
Off
Not connected or no connection established.
White
Connected and connection established.
Flashing white
Communication not active.
5
Port 2 active:
PROFINET and
service interface (CDI)
Off
Not connected or no connection established.
Yellow
Connected and connection established.
Flashing yellow
Communication not active.
Proline 500
Different LEDs in the transmitter provide information on the device status.
1
2
3
4
5
A0029629
1
Supply voltage
2
Device status
3
Flashing/network status
4
Port 1 active: PROFINET
5
Port 2 active: PROFINET and service interface (CDI)
LED
Color
Meaning
1
Supply voltage
Off
Supply voltage is off or too low.
Green
Supply voltage is ok.
2
Device status (normal
operation)
Off
Firmware error.
Green
Device status is ok.