Proline Promag W 500 PROFINET
Diagnostics and troubleshooting
Hauser
199
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
437 Configuration incompatible
1. Restart device
2. Contact service
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
438 Dataset
1. Check data set file
2. Check device configuration
3. Up- and download new configuration
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status
Quality
Uncertain
Quality substatus
Maintenance demanded
Coding (hex)
0x68 to 0x6B
Status signal
M
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
441 Current output 1 to n
1. Check process
2. Check current output settings
–
Measured variable status [from the factory]
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
S
Diagnostic behavior
Warning
1)
Diagnostic behavior can be changed. This causes the overall status of the measured variable to change.