Operation
Proline Promag W 500 PROFINET
166
Hauser
Options
Description
totalize
The totalizer is set to the defined start value from the
Preset value
parameter and
the totaling process is restarted.
Hold
Totalizing is stopped.
11.6.2 Function scope of the "Reset all totalizers" parameter
Options
Description
Cancel
No action is executed and the user exits the parameter.
Reset + totalize
Resets all totalizers to 0 and restarts the totaling process. This deletes all the flow
values previously totalized.
11.7 Showing data logging
The
Extended HistoROM
application package must be enabled in the device (order
option) for the
Data logging
submenu to appear. This contains all the parameters for the
measured value history.
Data logging is also available via:
• Plant Asset Management Tool FieldCare → 101.
• Web browser
Function range
• A total of 1000 measured values can be stored
• 4 logging channels
• Adjustable logging interval for data logging
• Displays the measured value trend for each logging channel in the form of a chart
0
/ ../XXXXXXXX
175.77
40.69 l/h
-100s
A0034352
• x-axis: depending on the number of channels selected displays 250 to 1000 measured
values of a process variable.
• y-axis: displays the approximate measured value span and constantly adapts this to the
ongoing measurement.
If the length of the logging interval or the assignment of the process variables to the
channels is changed, the content of the data logging is deleted.
Navigation
"Diagnostics" menu → Data logging
‣
Data logging
Assign channel 1
Assign channel 2
Assign channel 3