Commissioning
Proline Promag W 500 PROFINET
126
Hauser
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Operating mode
–
Define the output as a pulse,
frequency or switch output.
• Pulse
• Frequency
• Switch
Pulse
Terminal number
–
Shows the terminal numbers
used by the PFS output
module.
• Not used
• 24-25 (I/O 2)
• 22-23 (I/O 3)
• 20-21 (I/O 4)
–
Signal mode
–
Select the signal mode for the
PFS output.
• Passive
• Active
• Passive NAMUR
Passive
Assign pulse output 1 to n
The
Pulse
option is selected in
the
Operating mode
parameter parameter.
Select process variable for
pulse output.
• Off
• Volume flow
• Mass flow
• Corrected volume
flow
Off
Pulse scaling
The
Pulse
option is selected in
the
Operating mode
parameter (→ 125) and a
process variable is selected in
the
Assign pulse output
Enter quantity for measured
value at which a pulse is
output.
Positive floating
point number
Depends on country
and nominal
diameter
Pulse width
The
Pulse
option is selected in
the
Operating mode
parameter (→ 125) and a
process variable is selected in
the
Assign pulse output
Define time width of the
output pulse.
0.05 to 2
000 ms
100 ms
Failure mode
The
Pulse
option is selected in
the
Operating mode
parameter (→ 125) and a
process variable is selected in
the
Assign pulse output
Define output behavior in
alarm condition.
• Actual value
• No pulses
No pulses
Invert output signal
–
Invert the output signal.
• No
• Yes
No
Configuring the frequency output
Navigation
"Setup" menu → Pulse/frequency/switch output
‣
Pulse/frequency/switch output
1 to n
Operating mode
Terminal number
Signal mode
Assign frequency output