Proline Promag W 500 PROFINET
Electrical connection
Hauser
53
7.2.2
Connecting the transmitter
1
2 3
4
5
6
A0028200
1
Terminal connection for supply voltage
2
Terminal connection for signal transmission, input/output
3
Terminal connection for signal transmission, input/output
4
Terminal connection for connecting cable between sensor and transmitter
5
Terminal connection for signal transmission, input/output or terminal connection for network connection via
service interface (CDI-RJ45); optional: connection for external WLAN antenna
6
Protective earth (PE)
In addition to connecting the device via PROFINET and the available inputs/outputs,
additional connection options are also available:
• Integrate into a network via the service interface (CDI-RJ45) → 56.
• Integrate the device into a ring topology → 57.
Connecting the PROFINET connector
1.
3.
2.
4 x
TX 20
TX 20
22 mm
8.
24 mm
5.
6.
4.
≤
42 (1.6)
7.
A0033987
1. Loosen the 4 fixing screws on the housing cover.
2. Open the housing cover.
3. Fold open the terminal cover.
4. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
5. Strip the cable and cable ends and connect to the RJ45 connector.