Proline Promag W 500 PROFINET
Diagnostics and troubleshooting
Hauser
197
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
512 Sensor electronic (ISEM) faulty
1. Check ECC recovery time
2. Turn off ECC
• Conductivity
• Corrected conductivity
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status
Quality
Uncertain
Quality substatus
Maintenance demanded
Coding (hex)
0x68 to 0x6B
Status signal
F
Diagnostic behavior
Alarm
12.7.3 Diagnostic of configuration
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
330 Flash file invalid
1. Update firmware of device
2. Restart device
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
M
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
331 Firmware update failed
1. Update firmware of device
2. Restart device
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Warning