Diagnostics and troubleshooting
Proline Promag W 500 PROFINET
198
Hauser
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
410 Data transfer
1. Check connection
2. Retry data transfer
• Conductivity
• Corrected conductivity
• Measured values 1
• Measured values 2
• Measured values 3
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status
Quality
Bad
Quality substatus
Maintenance alarm
Coding (hex)
0x24 to 0x27
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
412 Processing download
Download active, please wait
• Conductivity
• Corrected conductivity
• Density
• Electronic temperature
• Flow velocity
• Mass flow
• Coil current shot time
• Reference electrode
potential against PE
• Noise
• Reference density
• Corrected volume flow
• Temperature
• Status
• Volume flow
Measured variable status
Quality
Uncertain
Quality substatus
Initial value
Coding (hex)
0x4C to 0x4F
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Influenced measured
variables
No.
Short text
431 Trim 1 to n
Carry out trim
–
Measured variable status
Quality
Good
Quality substatus
Ok
Coding (hex)
0x80 to 0x83
Status signal
C
Diagnostic behavior
Warning