Proline Promag W 500 PROFINET
System integration
Hauser
111
(input data)
4
Verification has failed
5
Verification performed successfully
6
Verification has not been performed
7
–
Start verification
(output data)
Verification control
A change in the status from 0 to 1 starts the verification
Data structure
Output data of the Heartbeat Verification module
Byte 1
Discrete Output
Input data of the Heartbeat Verification module
Byte 1
Byte 2
Discrete Input
Status
1)
Status coding → 111
9.3.3
Status coding
Status
Coding (hex)
Meaning
BAD - Maintenance alarm
0x24
A measured value is not available because a device
error has occurred.
BAD - Process related
0x28
A measured value is not available because the process
conditions are not within the device's technical
specification limits.
BAD - Function check
0x3C
A function check is active (e.g. cleaning or calibration)
UNCERTAIN - Initial value
0x4F
A pre-defined value is output until a correct measured
value is available again or until remedial measures
have been carried out that change this status.
UNCERTAIN -
Maintenance demanded
0x68
Signs of wear and tear have been detected on the
measuring device. Short-term maintenance is needed
to ensure that the measuring device remains
operational.
The measured value might be invalid. The use of the
measured value depends on the application.
UNCERTAIN - Process
related
0x78
The process conditions are not within the device's
technical specification limits. This could have a
negative impact on the quality and accuracy of the
measured value.
The use of the measured value depends on the
application.
GOOD - OK
0x80
No error has been diagnosed.
GOOD - Maintenance
demanded
0xA8
The measured value is valid.
It is highly advisable to service the device in the near
future.
GOOD - Function check
0xBC
The measured value is valid.
The measuring device is performing an internal
function check. The function check does not have any
noticeable effect on the process.