Electrical connection
Proline Promag W 500 PROFINET
48
Hauser
7.1.7
Preparing the connecting cable: Proline 500 – digital
When terminating the connecting cable, pay attention to the following points:
‣
For cables with fine-wire cores (stranded cables):
Fit the cores with ferrules.
Transmitter
Sensor
A
90 (3.54)*
70 (2.76)
50 (1.97)
10 (0.4)
8 (0.3)
B
1
A0029330
A
80 (3.15)
50 (1.97)
6 (0.24)
170 (6.7)*
20 (0.8)*
B
1
2
A0029443
Engineering unit mm (in)
A = Terminate the cable
B = Fit ferrules on cables with fine-wire cores (stranded cables)
1 = Red ferrules, 1.0 mm (0.04 in)
2 = White ferrules, 0.5 mm (0.02 in)
* = Stripping only for reinforced cables
7.1.8
Preparing the connecting cable: Proline 500
When terminating the connecting cable, pay attention to the following points:
1. In the case of the electrode cable:
Make sure that the ferrules do not touch the core shields on the sensor side.
Minimum distance = 1 mm (exception: green “GND” cable)
2. In the case of the coil current cable:
Insulate one core of the three-core cable at the level of the core reinforcement. You
only require two cores for the connection.
3. For cables with fine-wire cores (stranded cables):
Fit the cores with ferrules.