Technical data
Proline Promag W 500 PROFINET
232
Hauser
Supported connections
• 2 x AR (IO Controller AR)
• 1 x AR (IO-Supervisor Device AR connection allowed)
• 1 x Input CR (Communication Relation)
• 1 x Output CR (Communication Relation)
• 1 x Alarm CR (Communication Relation)
Configuration options for
measuring device
• DIP switches on the electronics module, for device name assignment (last
part)
• Manufacturer-specific software (FieldCare, DeviceCare)
• Web browser
• Device master file (GSD), can be read out via the integrated Web server of
the measuring device
Configuration of the
device name
• DIP switches on the electronics module, for device name assignment (last
part)
• DCP protocol
• Process Device Manager (PDM)
• Integrated Web server
Supported functions
• Identification & Maintenance
Simple device identification via:
• Control system
• Nameplate
• Measured value status
The process variables are communicated with a measured value status
• Blinking feature via the onsite display for simple device identification and
assignment
• Device operation via operating tools (e.g. FieldCare, DeviceCare, SIMATIC
PDM)
System integration
Information on system integration → 105 .
• Cyclic data transmission
• Overview and description of the modules
• Status coding
• Startup configuration
• Factory setting
16.5 Power supply
Terminal assignment
Device plugs available
Supply voltage
Order code for
"Power supply"
Terminal voltage
Frequency range
Option
D
DC24 V
±20%
–
Option
E
AC100 to 240 V –15…+10%
50/60 Hz, ±4 Hz
Option
I
DC24 V
±20%
–
AC100 to 240 V –15…+10%
50/60 Hz, ±4 Hz
Power consumption
Transmitter
Max. 10 W (active power)
switch-on current
Max. 36 A (<5 ms) as per NAMUR Recommendation NE 21