2-43
IM 701210-05E
Explanation of Functions
2
2.5
Recording in Recorder Mode (Realtime
Recording) (DL750P Only)
The DL750P is equipped with an A-4 size printer. If Chart Recorder mode is selected in
the RECORDER menu and waveform acquisition is started, T-Y waveforms (T-Y
waveform recording) or numeric values can be recorded in realtime. In addition, if X-Y
Recorder mode is selected and waveform acquisition is started and stopped, the X-Y
waveform over the interval in which waveform was acquired can be recorded (X-Y
waveform recording).
Chart Recorder Mode <Sections 9.2 to 9.6>
Print Style
In Chart Recorder mode, the print style can be selected from the following two styles.
T-Y Waveform Recording
In T-Y waveform recording, recording is performed at the specified chart speed (10
mm/h to 20 mm/s). The scale used in T-Y waveform recording is fixed to 10 mm per
division on the horizontal axis (time axis). The scale on the vertical axis (voltage axis)
can be set to 10 mm per division or [1 division of the 10 equally divided chart
recording zone] per division.The recording can also be stopped automatically after
recording a specified length (shot recording).
Numeric Value Recording
Numeric values can be recorded at the specified time interval (1 s to 60 min). Up to
16 channels can be recorded. If waveform acquisition is started on more than 16
channels, 16 channels are recorded in the following order: CH1 to CH6, DSP1 to
DSP6 (option), LOGIC A, and LOGIC B.
Acquisition Mode during Chart Recorder Mode
During Chart Recorder mode, the acquisition mode is set to envelope. The acquisition
mode cannot be changed with the ACQ key as with the normal operation.
However, the acquisition mode can be changed to normal with the MISC menu. Select
normal mode if you are using the acquired waveform as numeric data such as in an
analysis.
Selectable Trigger Modes in Chart Recorder Mode
The following four trigger modes can be selected.
Auto
Recording on the built-in printer starts at the same time measurement is started by
pressing the START/STOP key. Recording continues until measurement is stopped
by pressing the START/STOP key again. The maximum number of divisions of the
most recent data that can be saved (see page 2-45) is saved in the internal memory.
Measurement start
Record start
Measurement stop
Record stop
The maximum number of
divisions of the most recent data
is saved in the internal memory