6-7
IM 701210-05E
Triggering
6
Selectable Range of Hold Off Time
0 ns to 10 s (the initial value is 0 ns). The resolution is 100 ns.
Notes for Setting the Hold Off Time
• If you are setting the hold off time to 50 ms or greater, set the trigger mode to normal.
• When used with A->B(N) or A Delay B trigger, the hold off time applies only to
condition B.
• The hold off time is set to 0 ns for Period trigger, and this function is invalid.
6.4 Setting the Hold Off Time