7-15
IM 701210-05E
Acquisition and Display
7
• Displaying the List of Captured Waveforms: Show Map
The number of the acquired waveform and the trigger time are listed. You can also
select the desired waveform number to display the list.
• Send Mail
When turned ON, a mail is sent to the mail address specified in NetWork settings
each time a waveform is acquired. For details, see section 16.5, "Sending Periodic
Mail or Action Mail."
Mark
The
mark indicates the position of the sub waveform specified by Capture Num at the
top section of on the main waveform window. However, when waveforms are being
acquired in the Auto mode, the
mark is not displayed for sub waveform positions that
exceed the end of the main waveform.
Note
Key response is slowed down when waveform acquisition is in progress using the dual
capture function.
Notes When Using the Dual Capture Function
• The dual capture function can be used only when the trigger mode is set to Auto or
Log.
• The maximum record length of the main waveform varies depending on the model as
follows: The dual capture function cannot be used at record lengths that exceed the
maximum record length.
Standard model: 5 M
/M1 model: 10 M
/M2 model:
50 M
/M3 model: 100 M
• The dual capture function cannot be used if the acquisition mode of the main
waveform is Average.
• The dual capture function and the realtime recording function cannot be used
simultaneously.
• The dual capture mode cannot be changed while waveform acquisition is in progress.
• If you restart measurements after acquiring waveforms, all the waveforms that were
acquired before changing the mode are cleared.
• The dual capture function cannot be applied to X-Y waveforms.
• Saving/loading of the waveforms acquired using the dual capture function is
performed simultaneously for the main waveform and the sub waveform. For details
on the saving/loading of the waveform, see section 13.7, “Saving/Loading Waveform
Data.”
• The dual capture function cannot be used if the time base is set to external clock.
• Accumulate function cannot be used.
• Cursor measurements and automated measurement of waveform parameters are
suspended while waveforms are acquired using the dual capture function.
• Automated measurement of waveform parameters and computations cannot be
performed on the sub waveform.
• When the sub waveform is displayed, cursor measurements are made on the sub
waveform.
7.6 Using the Dual Capture Function