6.6
Setting the External Trigger (SIMPLE)
<For a description of this function, refer to page 2-20.>
Procedure
MEASURE CURSOR
RESET
SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP
DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV
TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8
9
DSP 2
6
5
4
1
0
2
3
ENTER
m
DSP 6
DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3
DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
1.
Press
SIMPLE/ENHANCED
.
2.
Press the
Select
soft key to select Simple.
Setting the Trigger Source
3.
Press the
Source
soft key. The trigger source selection menu appears.
4.
Use the
jog shuttle
and
SELECT
to select Ext.
Setting the Trigger Slope
5.
Press the
Slope
soft key to select or .
Setting the Hold Off
6.
Set the hold off time according to the procedures given in section 6.4.
DSP1 to DSP6 are optional.
Explanation
The external signal that is input through the TRIG IN terminal on the left side panel of the
DL750/DL750P can be used to generate triggers.
Note
For details on the TRIG IN terminal, see section 14.1.
Setting the Trigger Source
Select Ext.
Setting the Trigger Slope
Select how the trigger source is to cross the specified level for activating the trigger from
the following two choices.
:
Activated when the trigger source changes from below the trigger level to above the
trigger level (rising).
:
Activated when the trigger source changes from above the trigger level to below the
trigger level (falling).
Setting the Hold Off