5-40
IM 701210-05E
5.19 Setting the Frequency (Number of Rotations,
Period, Duty Cycle, Power Supply Frequency,
Pulse Width, Pulse Integration, and Velocity)
Measurement
<For a description of this function, refer to page 2-16.>
Procedure
1.
Press one of the keys from
CH1
to
CH16
to select the desired channel (select a
channel with the 701280(FREQ) installed).
Setting the Input
MEASURE CURSOR
RESET
SELECT
CH
ALL CH
MODE
POSITION
SIMPLE/ENHANCED
ACQ
START/STOP
SETUP
DISPLAY
ZOOM
DUAL
CAPTURE
HISTORY
MATH
1
CH
2
CH
3
CH
4
CH
5
CH
6
CH
7
CH
8
CH
9
CH
10
CH
11
CH
12
CH
13
CH
14
CH
15
CH
16
DELAY
MANUAL TRIG
V/DIV
TIME/DIV
TRIGGER
TRIG D
VERTICAL
HORIZONTAL
CAL
X-Y
SEARCH
7
DSP 1
8
9
DSP 2
6
5
4
1
0
2
3
ENTER
m
DSP 6
DSP 5
LOGIC A
LOGIC B
EVENT
DSP 3
DSP 4
EXP
GO/NO-GO
ACTION
FILE
SHIFT
Setting the Preset
2.
Press the
Input Setup
soft key. A setup dialog box opens.
3.
Use the
jog shuttle
and
SELECT
to select Preset according to the type of
sensor that is connected to the frequency module. When you select a preset
(Logic 5V to EM Pickup), the input is set to a value appropriate for the signal.
(Some items need to be manually entered.)
If you select User (user-defined), all the setup items can be set arbitrarily. For a
description of the setup items, see the next page.
Typical Applications and Signal Examples of Each Preset
Preset
Application and Signal Example
Logic 5V
5-V logic signal, 5-V output sensor, and sensor with TTL output
Logic 3V
3-V logic signal and 3-V output sensor
Logic 12V
12-V driven relay/sequence circuit and 12-V driven sensor
Logic 24V
24-V driven relay/sequence circuit and 24-V driven sensor
ZeroCross
Sensor/Encoder that outputs positive and negative voltages and sensor that
outputs sine waves
AC100V
When measuring a supply voltage of 100 VAC using the isolated probe (700929)
AC200V
When measuring a supply voltage of 200 VAC using the isolated probe (700929)
EM Pickup
Power-generating electromagnetic pickup
Pull -up 5V
Open-collector output sensor and contact circuit
The Pull Up setup menu appears only when
you set the preset to Pull-up 5V.