6-44
IM 701210-05E
Window Width
The realtime template is created by applying a tolerance width (window width) to the
averaged waveform. The selectable range of the window width varies depending on the
observed item as follows.
When observing voltage:
(V/div)
×
1/100 to (V/div)
×
10
When observing strain:
1
µ
STR to (measurement range)
×
2
Or, 0.0005 mV/V to (measurement range)
×
2
When measuring acceleration: 0.01 Unit to (Unit/div)
×
10
(The resolution is the same as the resolution of the trigger level. See section 6.5.)
For example, if Width is set to 2 V, a window width of
±
1 V is created with respect to the reference waveform.
Window width
1 V
1 V
Reference waveform
Hold Off
For details, see section 6.4, “Setting the Hold Off Time.”
Note
If a trigger waveform such as a serge is input immediately before the completion of the pre-
trigger, the trigger point may be delayed by up to 4 cycles.
6.17 Setting the Wave Window Trigger (ENHANCED)