5-51
IM 701210-05E
Horizontal and Vertical Axes
5
The following 10 input setup items are available.
• Voltage Range: V Range
Select the input voltage range (
±
FS) from the following:
• When the probe attenuation is set to 1:1
±
1 V,
±
2 V,
±
5 V,
±
10 V,
±
20 V, or
±
50 V (
±
FS)
• When the probe attenuation is set to 10:1
±
10 V,
±
20 V,
±
50 V,
±
100 V,
±
200 V, or
±
500 V (
±
FS)
• Coupling
Set the input coupling to DC or AC.
The description is the same as with the input coupling of other modules. See the
functional explanation given in “Input Coupling” on page 2-8 and the explanation given
in section 5.5.
• Probe Type
Select 1:1 or 10:1.
The description is the same as with the probe type of other modules. See the
functional explanation given in “Probe Type” on page 2-9 and the explanation given in
section 5.6.
• Bandwidth Limit
Set the bandwidth limit to 100 Hz, 1 kHz, 10 kHz, 100 kHz, or Full. You cannot select
Full when the preset is set to AC100V or AC200V.
• Threshold Level
Set the level within FS of the input voltage range. The resolution is a value
corresponding to 1% of FS.
• Hys (Hysteresis)
Set the hysteresis to
±
1%,
±
2.5%, or
±
5% of the FS of the input voltage range.
• Slope
Select (rising) or (falling).
• Chatter Elimination
Eliminates the chatter that occurs such when the contact input is turned ON/OFF.
The changes in the signal over the specified time can be discarded. Set the interval
in the range of 0 to 1000 ms (1-ms resolution). When 0 ms is selected, the chatter
elimination function is turned OFF.
Chattering elimination is effective against both the rising and falling edges.
• Pull Up
You can turn ON/OFF pull-up only when the preset is set to Pull-up 5V. Pull-up is not
possible in other presets.
When using pull-up, set the input voltage in the range of 0 V to 5 V. If a voltage exceeding
this range is applied, the internal protection circuit will cut off the pull-up circuit.
Internal equivalent circuit when using pull-up
+5V
4.7 k
Ω
990 k
Ω
10 k
Ω
5.19 Setting the Frequency (Number of Rotations, Period, Duty Cycle, Power Supply Frequency,
Pulse Width, Pulse Integration, and Velocity) Measurement