
5-39
IM 701210-05E
Horizontal and Vertical Axes
5
Setting the Bandwidth Limit and Waveform Label
For the setup procedure of the bandwidth limit, see section 5.7. For the setup procedure
of the waveform label, see section 8.10.
Explanation
Setting the Gain
Set the ratio of the output signal with respect to the input signal in the range of
×
0.1 to
×
100.
The default value is
×
1.
Setting the Sensitivity
Set the sensitivity of the acceleration sensor in the range of 0.10 mV/Unit to 2000.00 mV/
Unit.
Setting the Unit
Set the unit of the acceleration to be displayed on the screen using up to 4 characters.
The default setting is m/s
2
. Change the unit as necessary.
Up to 8 character strings assigned to the unit are stored. The stored strings can be
recalled using the soft key. In addition, the stored strings can be used on the
acceleration modules of other channels. This function is useful when you are assigning
the same unit to multiple channels. The stored strings are not cleared even when the
power is turned OFF.
Setting the Bias Current Supply to the Acceleration Sensor
When Bias is turned ON, a 4-mA bias current is supplied to the acceleration sensor. Do
not connect the acceleration sensor with Bias turned ON.
Setting the Bandwidth Limit
See section 5.7, “Setting the Bandwidth Limit.”
5.18 Setting the Acceleration Measurement