Chapter 3 - Installation
MNL000624.doc
Revision 0
06/25/2019
Chemical Equipment
© Versum Materials US, LLC. as of the revision and date shown. All rights reserved.
Confidential and Proprietary Data
Page 3
-
5
Versum Materials Confidential
Type 1 =
Equipment fully de-energized.
Type 2 =
Equipment is energized.
Live circuits are covered or insulated. Work is performed at a remote location to
preclude accidental shock.
3.4.2
Electrical Requirements
AC POWER
100 - 240 VAC, 100 W @ 50 - 60 Hz; Single-Phase, 3 wires; Neutral solidly
grounded, without optional degasser and pump
125 VAC,1000 W @ 50 or 60 Hz; Single-Phase, 3 wires; Neutral solidly grounded,
with optional degasser pump. Short circuit current rating (SCCR) is 10kA.
OSHA standards require customer to install lockout-type circuit breaker for AC
power. Versum Materials, Inc. recommends an over current protection of 10,000
Ampere Interrupt Capacity (AIC) be provided in close proximity to the unit.
WARNING
WARNING: Electrical
connections should be made
by a qualified electrician. AC
power junction box requires
Lock-out Tag-out ON/OFF
capability. ½’ AC power
conduit hole exits top of
cabinet.
WARNUNG: Elektrische
Anschlüsse sollten nur durch
qualifizierte Elektriker
hergestellt werden.
Wechselstrom-Verteilerkästen
müssen einen EIN/AUS-
Schalter zur Verriegelung
bzw. Isolierung besitzen.
Oben am Schrank wird ein ca.
15 cm (0,5 Fuß) langes
flexibles Schutzrohr für
Wechselstromleitungen
herausgeführt.
AVERTISSEMENT: Les
connexions électriques
doivent être réalisées par un
électricien qualifié. La boîte de
dérivation de courant alternatif
requiert une capacité de
connexion et déconnection
"ON/OFF". 15 cm (0,5 pieds)
du tube souple de courant
alternatif ressort du haut de la
boîte.