Table 6-5: Tasks available on the Quantum system (continued)
Task Name
Adherent Cells
Suspension
Cells
Purpose
Page
Load and Attach Tasks
Load Cells With
Circulation
Yes
Yes
Allows you to load the cells from the cell inlet
bag into the IC circulation loop until the bag is
empty, and uses IC circulation to distribute the
cells.
Load Cells Without
Circulation
Yes
Yes
Allows you to load the cells from the cell inlet
bag into the bioreactor via the IC circulation
loop until the bag is empty.
Attach Cells
Yes
No
Allows adherent cells to attach to the bioreactor
membrane while allowing flow in the
EC circulation loop.
Load Cells With
Uniform Suspension
Yes
Yes
Allows you to load the cells into the bioreactor
from the cell inlet bag until the bag is empty and
to use IC circulation to create uniform cell
suspension.
Feed and Add Tasks
Feed Cells With
Ultrafiltration
Yes
Yes
Allows you to continuously add fluid at a low
flow rate to the IC circulation loop and/or the
EC circulation loop. This task uses ultrafiltration
to remove excess IC fluid.
Feed Cells
Yes
No
Allows you to continuously add fluid at a low
flow rate to the IC circulation loop and/or the
EC circulation loop. There are several outlet
settings that you can use to remove the fluid
that is added to the system during this task.
Add Bolus
Yes
Yes
Allows you to quickly add a selected volume of
reagent into the IC circulation loop; you can also
add a bolus into the EC circulation loop at the
same time.
Add Bag Contents
Yes
Yes
Allows you to load a fluid, such as a reagent,
into the IC circulation loop until the bag that
contains the fluid is empty.
Add Fluid
Continuously
Yes
No
Allows you to continuously add fluid at a low
flow rate to the IC circulation loop and/or the
EC circulation loop. There are several outlet
settings that you can use to remove the fluid
that is added to the system during this task.
Add Fluid
Continuously With
Ultrafiltration
Yes
Yes
Allows you to continuously add fluid at a low
flow rate to the IC circulation loop and/or the
EC circulation loop. This task uses ultrafiltration
to remove excess IC fluid.
Table of Tasks
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
6-21
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...