Add Fluid Continuously
The purpose of this task is to continuously add fluid at a low flow rate to the IC circulation loop and/or
the EC circulation loop. There are several outlet settings that you can use to remove the fluid added to
the system during this task.
Table 11-10 describes the types of solution that are needed to attach to each line when performing Add
Fluid Continuously. These solutions and corresponding volumes are based on the factory default
settings for this task.
Table 11-10: Solutions for Add Fluid Continuously
Line
Solution Attached to the Line
Required Volume (estimate
based on factory default values)
Cell
None
N/A
Reagent
None
N/A
IC Media
Media with protein
6 mL/hour
Wash
None
N/A
EC Media
None
N/A
Add Fluid Continuously pathway:
Task
>
Feed and Add
>
Add Fluid Continuously
1.
Confirm the values for each setting for Add Fluid Continuously shown in Table 11-11.
Table 11-11: Settings for Add Fluid Continuously
Setting
Factory Default Values
Setting Options
IC Inlet
IC Media
Cell, Reagent, IC Media, Wash, EC Media, None
IC Inlet Rate
0.1 mL/min
0 to 10 mL/min
IC Circulation Rate
20 mL/min
−100 to 100 mL/min
The system automatically calculates this rate based on the IC Inlet Rate
for this task.
EC Inlet
None
Reagent, IC Media, Wash, EC Media, None
EC Inlet Rate
0 mL/min
0 to 10 mL/min
EC Circulation Rate
30 mL/min
10 to 300 mL/min
Outlet
IC Outlet
EC Outlet, IC Outlet, Synchronization
Rocker
Stationary (0°)
In Motion (−180° to 270°, 0 to 3600 sec), Stationary
(−180° to 270°)
Feed and Add Tasks
11-10
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...