Stationary
The Stationary rocker setting includes a single value, which denotes the fixed position the bioreactor
holds for the duration of a step of a task. The position is a certain number of degrees from the home
position of 0°, and it must fall within the range of −180° to 270°. The following example shows how this
operator's manual denotes Stationary as the rocker setting:
Stationary (45°)
Selecting Stationary as the Rocker Setting
The following instructions explain how to select Stationary as the rocker setting for a step of a task. For
information about using the data entry pad, see “Using the Data Entry Pad” on page 3-21.
From the Setup screen:
1.
Touch
Rocker
.
The Setup: Rocker window opens.
2.
Touch
Stationary
.
The Position field appears, and
Stationary
changes to black.
3.
Touch the Position field.
A data entry pad appears.
4.
Use the data entry pad to enter the desired number of degrees from the home position of 0°.
5.
Touch
Enter
on the data entry pad.
The data entry pad closes, and the Position field displays the position you entered.
6.
Touch
Confirm
.
The Setup: Rocker window closes.
Rocker
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
6-13
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...