Table 9-8: Settings for IC EC Washout Through Membrane (continued)
Setting
Factory Default
Setting Options
EC Inlet Rate
148 mL/min
N/A
The system automatically calculates this rate based on the number of
IC Volumes and EC Volumes used for the Exchange stop condition for this
task.
EC Circulation Rate
−1.7 mL/min
N/A
The system automatically calculates this rate based on the EC Inlet Rate
for this task.
Outlet
EC Outlet
N/A
Rocker
In Motion (−90°, 180°, 1 sec)
In Motion (−180° to 270°, 0 to 3600 sec), Stationary
(−180° to 270°)
Stop Condition
Exchange (1.0 IC Volumes)
(1.0 EC Volumes)
Exchange (0.5 to 5.0 IC Volumes)
(0.5 to 5.0 EC Volumes)
2.
Touch
Start
.
3.
Check the status line to confirm that the task has started.
When the system completes the task, the home screen appears, the system changes to an idle state,
and the status line displays Idle.
Washout Tasks
9-8
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...