Table 10-2: Step 1 Settings for Load Cells With Circulation (continued)
Setting
Factory Default
Setting Options
EC Inlet Rate
0 mL/min
N/A
EC Circulation Rate
30 mL/min
10 to 300 mL/min
Outlet
EC Outlet
N/A
Rocker
In Motion (−90°, 180°, 1 sec)
In Motion (−180° to 270°, 0 to 3600 sec), Stationary
(−180° to 270°)
Stop Condition
Empty Bag
N/A
2.
Confirm the settings for step 2 shown in Table 10-3.
Table 10-3: Step 2 Settings for Load Cells With Circulation
Setting
Factory Default
Setting Options
IC Inlet
IC Media
Reagent, IC Media, Wash, EC Media
IC Inlet Rate
50 mL/min
Same as step 1
IC Circulation Rate
139 mL/min
Same as step 1
EC Inlet
None
N/A
EC Inlet Rate
0 mL/min
N/A
EC Circulation Rate
30 mL/min
Same as step 1
Outlet
EC Outlet
N/A
Rocker
In Motion (−90°, 180°, 1 sec)
Same as step 1
Stop Condition
IC Volume (22 mL)
IC Volume (1 to 100 mL)
3.
Confirm the settings for step 3 shown in Table 10-4.
Table 10-4: Step 3 Settings for Load Cells With Circulation
Setting
Factory Default
Setting Options
IC Inlet
IC Media
Reagent, IC Media, Wash, EC Media
IC Inlet Rate
50 mL/min
Same as step 1
IC Circulation Rate
−41 mL/min
N/A
The system automatically calculates this rate based on the IC Inlet Rate
for this step.
Load Cells With Circulation
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
10-3
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...