IC EC Washout
The purpose of this task is to replace the fluid in both the IC circulation loop and the EC circulation
loop. The replacement volume is specified by the number of IC Volumes and EC Volumes exchanged.
Table 9-5 describes the types of solution that are needed to attach to each line when performing IC EC
Washout. These solutions and corresponding volumes are based on the factory default settings for this
task.
Table 9-5: Solutions for IC EC Washout
Line
Solution Attached to the
Line
Required Volume (estimate based on factory
default settings)
Cell
None
N/A
Reagent
None
N/A
IC Media
Media with protein
0.6 L
Wash
None
N/A
EC Media
Media without protein
0.9 L
IC EC Washout pathway:
Task
>
Washout
>
IC EC Washout
1.
Confirm the settings for IC EC Washout shown in Table 9-6.
Table 9-6: Settings for IC EC Washout
Setting
Factory Default
Setting Options
IC Inlet
IC Media
Reagent, IC Media, Wash, EC Media
IC Inlet Rate
100 mL/min
N/A
The system automatically calculates this rate based on the EC Inlet Rate
and the number of IC Volumes and EC Volumes used for the Exchange
stop condition for this task.
IC Circulation Rate
−17 mL/min
N/A
The system automatically calculates this rate based on the IC Inlet Rate
for this task.
EC Inlet
EC Media
Reagent, IC Media, Wash, EC Media
EC Inlet Rate
148 mL/min
N/A
The system automatically calculates this rate based on the number of
IC Volumes and EC Volumes used for the Exchange stop condition for this
task.
IC EC Washout
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
9-5
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...