High Density Washout
The purpose of this task is to remove any non
‑
adhered cells from the bioreactor after loading 25 mL to
62 mL of bone marrow. The types of cells removed include red blood cells, platelets, and non-adherent
mononuclear cells.
This task is also a useful washout task for any occasion when the bioreactor is packed with a large number
of cells. When loading 10 mL or less of bone marrow, the Rapid IC Washout task is recommended.
When loading between 10 mL and 25 mL of bone marrow, no specific guidelines exist; using this task in
such cases should not cause problems but may not be necessary.
Table 9-1 describes the types of solution that are needed to attach to each line when performing High
Density Washout. These solutions and corresponding volumes are based on the factory default settings
for this task.
Table 9-1: Solutions for High Density Washout
Line
Solution Attached to the
Line
Required Volume (estimate based on factory
default settings)
Cell
None
N/A
Reagent
None
N/A
IC Media
Media with protein
Approximately 0.5 L
Wash
None
N/A
EC Media
Media without protein
Approximately 0.5 L
High Density Washout pathway:
Task
>
Washout
>
High Density Washout
1.
Confirm the settings for High Density Washout shown in Table 9-2.
Table 9-2: Settings for High Density Washout
Setting
Factory Default
Setting Options
IC Inlet
IC Media
IC Media, Wash, EC Media
Maximum IC Inlet
Rate
100 mL/min
80 to 200 mL/min
EC Inlet
EC Media
IC Media, Wash, EC Media
EC Circulation Rate
30 mL/min
10 to 300 mL/min
Stop Condition
Inlet Volume (1000 mL)
Inlet Volume (400 to 4000 mL)
2.
Touch
Start
.
3.
Check the status line to confirm that the task has started.
When the system completes the task, the home screen appears, the system changes to an idle state,
and the status line displays Idle.
Washout Tasks
9-2
Quantum
®
Cell Expansion System • Operator's Manual for Software Version 2.1
Summary of Contents for Quantum Cell Expansion System
Page 1: ...Quantum Cell Expansion System SOFTWARE VERSION 2 1 Operator s Manual...
Page 9: ...vi Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 10: ...1 Preface Quantum Cell Expansion System Operator s Manual for Software Version 2 1 1 1...
Page 19: ...Preface 1 10 Quantum Cell Expansion System Operator s Manual for Software Version 2 1...
Page 20: ...2 Introduction Quantum Cell Expansion System Operator s Manual for Software Version 2 1 2 1...
Page 38: ...3 Touch Screen Quantum Cell Expansion System Operator s Manual for Software Version 2 1 3 1...
Page 60: ...4 Troubleshooting Quantum Cell Expansion System Operator s Manual for Software Version 2 1 4 1...
Page 120: ...9 Washout Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 9 1...
Page 166: ...13 Custom Tasks Quantum Cell Expansion System Operator s Manual for Software Version 2 1 13 1...
Page 172: ...15 Reports Quantum Cell Expansion System Operator s Manual for Software Version 2 1 15 1...